Počet záznamů: 1  

Iodine Absorption Cells Purity Testing

  1. 1.
    SYSNO ASEP0468639
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevIodine Absorption Cells Purity Testing
    Tvůrce(i) Hrabina, Jan (UPT-D) RID, ORCID, SAI
    Zucco, M. (IT)
    Philippe, Ch. (FR)
    Pham, Minh Tuan (UPT-D) RID, ORCID, SAI
    Holá, Miroslava (UPT-D) RID, ORCID, SAI
    Acef, O. (FR)
    Lazar, Josef (UPT-D) RID, ORCID, SAI
    Číp, Ondřej (UPT-D) RID, SAI, ORCID
    Celkový počet autorů8
    Číslo článku17010102
    Zdroj.dok.Sensors. - : MDPI
    Roč. 17, č. 1 (2017), s. 1-13
    Poč.str.13 s.
    Forma vydáníTištěná - P
    Jazyk dok.eng - angličtina
    Země vyd.CH - Švýcarsko
    Klíč. slovaiodine cells ; absorption spectroscopy ; laser spectroscopy ; laser standards ; frequency stability
    Vědní obor RIVBH - Optika, masery a lasery
    Obor OECDOptics (including laser optics and quantum optics)
    CEPGA15-18430S GA ČR - Grantová agentura ČR
    LO1212 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    ED0017/01/01 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    Institucionální podporaUPT-D - RVO:68081731
    UT WOS000393021000102
    DOI10.3390/s17010102
    AnotaceThis article deals with the evaluation of the chemical purity of iodine-filled absorption cells and the optical frequency references used for the frequency locking of laser standards. We summarize the recent trends and progress in absorption cell technology and we focus on methods for iodine cell purity testing. We compare two independent experimental systems based on the laser-induced fluorescence method, showing an improvement of measurement uncertainty
    by introducing a compensation system reducing unwanted influences. We show the advantages
    of this technique, which is relatively simple and does not require extensive hardware equipment. As an alternative to the traditionally used methods we propose an approach of hyperfine transitions’ spectral linewidth measurement. The key characteristic of this method is demonstrated on a set of testing iodine cells. The relationship between laser-induced fluorescence and transition linewidth methods will be presented as well as a summary of the advantages and disadvantages of the proposed technique (in comparison with traditional measurement approaches).