Počet záznamů: 1  

High resolution finite element modeling of cemented bone-implant interface using X-ray microtomography

  1. 1.
    SYSNO ASEP0353055
    Druh ASEPC - Konferenční příspěvek (mezinárodní konf.)
    Zařazení RIVD - Článek ve sborníku
    NázevHigh resolution finite element modeling of cemented bone-implant interface using X-ray microtomography
    Tvůrce(i) Kytýř, Daniel (UTAM-F) SAI, RID, ORCID
    Jiroušek, Ondřej (UTAM-F) RID, SAI
    Zlámal, Petr (UTAM-F) RID, SAI, ORCID
    Pokorný, D. (CZ)
    Dammer, J. (CZ)
    Celkový počet autorů5
    Zdroj.dok.Proceedings of 9th International Symposium on Computer methods in Biomechanics and Biomedical Engineering. - Cardiff : ARUP, 2010 / Middleton John - ISBN 978-0-9562121-3-9
    Rozsah strans. 467-471
    Poč.str.5 s.
    Forma vydáníCD-ROM - CD-ROM
    AkceInternational symposium on Computer methods in biomechanics and biomedial engineering /9./
    Datum konání24.02.2010-27.02.2010
    Místo konáníValencia
    ZeměES - Španělsko
    Typ akceWRD
    Jazyk dok.eng - angličtina
    Země vyd.GB - Velká Británie
    Klíč. slovabone-cement interface ; computed tomography ; high-resolution models
    Vědní obor RIVFI - Traumatologie, ortopedie
    CEPGAP105/10/2305 GA ČR - Grantová agentura ČR
    CEZAV0Z20710524 - UTAM-F (2005-2011)
    AnotaceThe aim of the research was to investigate the cemented bone-implant interface behavior. To determine the degradation caused by mechanical loading both numerical and experimental approach have been used. The main problems (cement layer degradation and bone-cement interface debonding) during physiological loading conditions have been investigated using a custom hip simulator. The experimental setup was designed to allow cyclic loading of a sample of pelvic bone with implanted cemented acetabular component. The hip contact force of required direction and magnitude was applied to the implant using a spherical femoral component head. The most unfavorable activity (downstairs walking) was simulated in two million cycles with 4Hz frequency. The process of damage accumulation in cement fixation was monitored by repeated scanning using high resolution micro-focus X-ray Computed Tomography (microCT).
    PracovištěÚstav teoretické a aplikované mechaniky
    KontaktKulawiecová Kateřina, kulawiecova@itam.cas.cz, Tel.: 225 443 285
    Rok sběru2011
Počet záznamů: 1  