Počet záznamů: 1  

Narrow-selection bandwith of femtosecond laser comb with application to changes in optical path distance

  1. 1.
    SYSNO ASEP0352220
    Druh ASEPC - Konferenční příspěvek (mezinárodní konf.)
    Zařazení RIVD - Článek ve sborníku
    NázevNarrow-selection bandwith of femtosecond laser comb with application to changes in optical path distance
    Tvůrce(i) Šmíd, Radek (UPT-D) RID, SAI
    Ježek, Jan (UPT-D) RID, ORCID, SAI
    Buchta, Zdeněk (UPT-D) RID, SAI, ORCID
    Čížek, Martin (UPT-D) RID, ORCID, SAI
    Lazar, Josef (UPT-D) RID, ORCID, SAI
    Číp, Ondřej (UPT-D) RID, SAI, ORCID
    Celkový počet autorů6
    Zdroj.dok.Optical Micro- and Nanometrology III (Proceedings of SPIE Vol. 7718). - Bellingham : SPIE, 2010 - ISBN 978-0-8194-8191-7
    Rozsah stran771818: 1-8
    Poč.str.8 s.
    AkceOptical Micro- and Nanometrology /3./
    Datum konání13.04.2010
    Místo konáníBrussels
    ZeměBE - Belgie
    Typ akceWRD
    Jazyk dok.eng - angličtina
    Země vyd.US - Spojené státy americké
    Klíč. slovainterferometry ; cavity ; optical frequency synthesis
    Vědní obor RIVBH - Optika, masery a lasery
    CEZAV0Z20650511 - UPT-D (2005-2011)
    UT WOS000285182000034
    AnotaceIn this contribution we propose a scheme of Fabry-Perot interferometer measuring the absolute distance in atmosferic conditions using a femtosecond laser comb. The spacing of mirrors of the Fabry-Perot interferometer represents the length standard referenced to stable optical frequency of the femtosecond mode-locked laser. With the help of highly selective optical filter it is possible to get only a few of separate spectral components. By tuning and locking of the Fabry-Perot cavity to a selected single component it is possible to get a mechanical length standard with the uncertainty of the repetition frequency of the femtosecond laser. If the interferometer measures distance in atmospheric conditions, the absolute value of the laser wavelength fluctuates with a refractive index of air. Compairing the measurement in evacuated chamber with measurement in ambient air leads to enhanced precision in measurement of refrective index of air.
    PracovištěÚstav přístrojové techniky
    KontaktMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Rok sběru2011
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.