Počet záznamů: 1  

Speciation analysis of arsenic by selective hydride generation- cryotrapping-atomic fluorescence spectrometry with flame-in-gas- shield atomizer: Achieving extremely low detection limits with inexpensive instrumentation

  1. 1.
    0434308 - ÚIACH 2015 RIV US eng J - Článek v odborném periodiku
    Musil, Stanislav - Matoušek, Tomáš - Currier, J. M. - Stýblo, M. - Dědina, Jiří
    Speciation analysis of arsenic by selective hydride generation- cryotrapping-atomic fluorescence spectrometry with flame-in-gas- shield atomizer: Achieving extremely low detection limits with inexpensive instrumentation.
    Analytical Chemistry. Roč. 86, č. 20 (2014), s. 10422-10428. ISSN 0003-2700. E-ISSN 1520-6882
    Grant CEP: GA ČR GA14-23532S; GA MŠMT LH12040
    Institucionální podpora: RVO:68081715
    Klíčová slova: speciation analysis of arsenic * selective hydride generation * flame-in-gas-shield atomizer * cryotrapping-atomic fluorescence spectrometry
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 5.636, rok: 2014

    This work describes the method of a selective hydride generation-cryotrapping (HG-CT) coupled to an extremely sensitive but simple in-house assembled and designed atomic fluorescence spectrometry (AFS) instrument for determination of toxicologically important As species. Here, an advanced flame-in-gas-shield atomizer (FIGS) was interfaced to HG-CT and its performance was compared to a standard miniature diffusion flame (MDF) atomizer.