Simulation-based design of a microfabricated pneumatic electrospray nebulizer

  1. 1.
    0434464 - ÚIACH 2015 RIV DE eng J - Článek v odborném periodiku
    Járvás, G. - Grym, Jakub - Foret, František - Guttman, A.
    Simulation-based design of a microfabricated pneumatic electrospray nebulizer.
    Electrophoresis. Roč. 36, č. 3 (2015), s. 386-392. ISSN 0173-0835. E-ISSN 1522-2683
    Grant CEP: GA ČR(CZ) GBP206/12/G014
    Institucionální podpora: RVO:68081715
    Klíčová slova: CFD * microfabrication * modeling * electrospray * nebulizer
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 2.482, rok: 2015

    A microfabricated pneumatic electrospray nebulizer has been developed and evaluated using computer simulations and experimental measurements of the mass spectrometry signals. The microdevice under development is designed for electrospray mass spectrometry interfacing without the need to fabricate an electrospray needle and can be used as a disposable or an integral part of a reusable system. The design of the chip layout was supported by computational fluid dynamics simulations. The tested microdevices were fabricated in glass using conventional photolithography, followed by wet chemical etching and thermal bonding. The performance of the microfabricated nebulizer was evaluated by means of time-of-flight mass spectrometry with a peptide mixture. It was demonstrated that the nebulizer, operating at supersonic speed of the nebulizing gas, produced very stable nanospray (900 nl/min) as documented by less than 0.1 % (standard error) fluctuation in total mass spectrometric signal intensity.
    Trvalý link: http://hdl.handle.net/11104/0238535

     
    Název souboruStaženoVelikostKomentářVerzePřístup
    0434464_IR.pdf101 MBAutorský postprintvyžádat
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.