Počet záznamů: 1  

Transmutation detectors

  1. 1.
    0368640 - ÚJF 2013 RIV NL eng J - Článek v odborném periodiku
    Viererbl, L. - Lahodová, Z. - Klupák, V. - Sus, F. - Kučera, Jan - Kůs, P. - Marek, M.
    Transmutation detectors.
    Nuclear Instruments & Methods in Physics Research Section A. Roč. 632, č. 1 (2011), s. 109-111. ISSN 0168-9002. E-ISSN 1872-9576
    Výzkumný záměr: CEZ:AV0Z10480505
    Klíčová slova: Transmutation detector * Activation method * Neutron detector * Neutron fluence
    Kód oboru RIV: BG - Jaderná, atomová a mol. fyzika, urychlovače
    Impakt faktor: 1.207, rok: 2011

    We have designed a new type of detectors, called transmutation detectors, which can be used primarily for neutron fluence measurement. The transmutation detector method differs from the commonly used activation detector method in evaluation of detector response after irradiation. Instead of radionuclide activity measurement using radiometric methods, the concentration of stable non-gaseous nuclides generated by transmutation in the detector is measured using analytical methods like mass spectrometry.
    Trvalý link: http://hdl.handle.net/11104/0202932

     
     