Wet effluent diffusion denuder technique and determination of volatile organic compounds in air. I. Oxo compounds (alcohols and ketones)

  1. 1.
    0162262 - UIACH-O 20010010 RIV NL eng J - Článek v odborném periodiku
    Sklenská, Jana - Pařízek, Petr - Večeřa, Zbyněk
    Wet effluent diffusion denuder technique and determination of volatile organic compounds in air. I. Oxo compounds (alcohols and ketones).
    Journal of Chromatography A. Roč. 918, č. 1 (2001), s. 153-158. ISSN 0021-9673. E-ISSN 1873-3778
    Grant CEP: GA ČR GA203/98/0943
    Výzkumný záměr: CEZ:AV0Z4031919
    Klíčová slova: diffusion denuders * alcohols * air analysis
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 2.793, rok: 2001

    The wet effluent diffusion denuder (WEDD) preconcentration technique for the determination of methanol, ethanol, 1-propanol, 1-butanol, 1-pentanol, acetone, methyl ethyl ketone, diethyl ketone and methyl n-propyl ketone in air are discussed. The compounds are continuously collected into a thin film of deionized water flowing down the inner wall of the cylindrical wet effluent diffusion denuder. The concentrate is analysed by gas chromatography. Detection limits of alcohols and ketones are as low as 1mikrog/l (GC-flame ionization detector) and/or 1 ng/l (GC-MS). This technique could be applicable for the continuous monitoring of ppb (v/v) level of both and ketones in the air.