Počet záznamů: 1  

Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma

  1. 1.
    0134065 - FZU-D 20020354 RIV PL eng J - Článek v odborném periodiku
    Krása, Josef - Pietriková, Monika - Juha, Libor - Cejnarová, Andrea
    Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma.
    Nukleonika. Roč. 46, Suppl. (2001), s. S49-S51. ISSN 0029-5922. E-ISSN 1508-5791
    Grant CEP: GA MŠMT LA 055
    Výzkumný záměr: CEZ:AV0Z1010921
    Klíčová slova: dosimetry of photons * thermoluminescent dosimeters * X-ray diagnostics
    Kód oboru RIV: BL - Fyzika plazmatu a výboje v plynech
    Impakt faktor: 0.360, rok: 2001

    Responses of thermoluminescet dosimeters of CaF2:Dy, CaF2:Mn, and LiF:Mg,Cu,P to the deposited energy of monochromatic 1.75-3.5 keV radiation were measured with respect to an increase in the local dose due to the attenuation length that was much shorter than the TLD thickness. The responses of TLDs to X-ray emission from the laser-produced aluminium plasma were found to be independent of the dose rate within a wide range.