Počet záznamů: 1
Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma
- 1.0134065 - FZU-D 20020354 RIV PL eng J - Článek v odborném periodiku
Krása, Josef - Pietriková, Monika - Juha, Libor - Cejnarová, Andrea
Limitations of thermoluminescent dosimetry in soft X-ray diagnostics of pulsed plasma.
Nukleonika. Roč. 46, Suppl. (2001), s. S49-S51. ISSN 0029-5922. E-ISSN 1508-5791
Grant CEP: GA MŠMT LA 055
Výzkumný záměr: CEZ:AV0Z1010921
Klíčová slova: dosimetry of photons * thermoluminescent dosimeters * X-ray diagnostics
Kód oboru RIV: BL - Fyzika plazmatu a výboje v plynech
Impakt faktor: 0.360, rok: 2001
Responses of thermoluminescet dosimeters of CaF2:Dy, CaF2:Mn, and LiF:Mg,Cu,P to the deposited energy of monochromatic 1.75-3.5 keV radiation were measured with respect to an increase in the local dose due to the attenuation length that was much shorter than the TLD thickness. The responses of TLDs to X-ray emission from the laser-produced aluminium plasma were found to be independent of the dose rate within a wide range.
Trvalý link: http://hdl.handle.net/11104/0032002