Počet záznamů: 1  

Suppression of the non-linear background in a multimode fibre CARS endoscope

  1. 1.
    SYSNO ASEP0553554
    Druh ASEPJ - Článek v odborném periodiku
    Zařazení RIVJ - Článek v odborném periodiku
    Poddruh JČlánek ve WOS
    NázevSuppression of the non-linear background in a multimode fibre CARS endoscope
    Tvůrce(i) Pikálek, Tomáš (UPT-D) RID, ORCID, SAI
    Stibůrek, Miroslav (UPT-D)
    Simpson, Stephen Hugh (UPT-D) RID, SAI
    Čižmár, Tomáš (UPT-D) RID, ORCID, SAI
    Traegaardh, Johanna (UPT-D) RID, ORCID, SAI
    Celkový počet autorů5
    Číslo článku450375
    Zdroj.dok.Biomedical Optics Express. - : Optical Society of America - ISSN 2156-7085
    Roč. 13, č. 2 (2022), s. 862-874
    Poč.str.13 s.
    Forma vydáníOnline - E
    Jazyk dok.eng - angličtina
    Země vyd.US - Spojené státy americké
    Klíč. slovaresolution ; microscopy ; probe ; myelin ; light
    Vědní obor RIVBH - Optika, masery a lasery
    Obor OECDOptics (including laser optics and quantum optics)
    CEPEF15_003/0000476 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    EF16_013/0001775 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    EF18_046/0016045 GA MŠMT - Ministerstvo školství, mládeže a tělovýchovy
    Výzkumná infrastrukturaCzech-BioImaging II - 90129 - Ústav molekulární genetiky AV ČR, v. v. i.
    Způsob publikováníOpen access
    Institucionální podporaUPT-D - RVO:68081731
    UT WOS000750935000001
    EID SCOPUS85123415709
    DOI10.1364/BOE.450375
    AnotaceMultimode fibres show great potential for use as miniature endoscopes for imaging deep in tissue with minimal damage. When used for coherent anti-Stokes Raman scattering (CARS) microscopy with femtosecond excitation sources, a high band-width probe is required to efficiently focus the broadband laser pulses at the sample plane. Although graded-index (GRIN) fibres have a large bandwidth, it is accompanied by a strong background signal from four-wave mixing and other non-linear processes occurring inside the fibre. We demonstrate that using a composite probe consisting of a GRIN fibre with a spliced on step-index fibre reduces the intensity of the non-linear background by more than one order of magnitude without significantly decreasing the focusing performance of the probe. Using this composite probe we acquire CARS images of biologically relevant tissue such as myelinated axons in the brain with good contrast.
    PracovištěÚstav přístrojové techniky
    KontaktMartina Šillerová, sillerova@ISIBrno.Cz, Tel.: 541 514 178
    Rok sběru2023
    Elektronická adresahttps://opg.optica.org/boe/fulltext.cfm?uri=boe-13-2-862&id=468619