On-line, real time monitoring of exhaled trace gases by SIFT-MS in the perioperative setting: a feasibility study

  1. 1.
    0363284 - ÚFCH JH 2012 RIV GB eng J - Článek v odborném periodiku
    Boshier, P. R. - Cushnir, J. R. - Mistry, V. - Knaggs, A. - Španěl, Patrik - Smith, D. - Hanna, G. B.
    On-line, real time monitoring of exhaled trace gases by SIFT-MS in the perioperative setting: a feasibility study.
    Analyst. Roč. 136, č. 16 (2011), s. 3233-3237. ISSN 0003-2654. E-ISSN 1364-5528
    Výzkumný záměr: CEZ:AV0Z40400503
    Klíčová slova: SIFT-MS * breath analysis
    Kód oboru RIV: CF - Fyzikální chemie a teoretická chemie
    Impakt faktor: 4.230, rok: 2011

    A study is described of the first on line, real time analyses of the exhaled breath of five anaesthetized patients during the complete perioperative periods of laparoscopic surgery. These breath analyses were achieved using a selected ion flow tube, SIFT-MS, instrument, located in the operating theatre at an acceptable distance from the operating table, and coupled to the endotracheal tube in the ventilation circuit via a 5 metre long capillary tube. Thus, inhalation/exhalation breathing cycles, set to be at a frequency of 10 per minute, were sampled continuously for water vapour, the metabolites acetone and isoprene and the propofol used to induce anaesthesia for each operating period that ranged from 20 min (shortest) to 80 min (longest). Whilst there was some loss of water vapour along the long sampling line, the concentrations of the other trace compounds were not diminished.
    Trvalý link: http://hdl.handle.net/11104/0199269

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.