Počet záznamů: 1  

Effect of complexones and tensides on selectivity of nitrogen dioxide determination in air with a chemiluminescence aerosol detector

  1. 1.
    0162177 - UIACH-O 200020 RIV NL eng J - Článek v odborném periodiku
    Mikuška, Pavel - Večeřa, Zbyněk
    Effect of complexones and tensides on selectivity of nitrogen dioxide determination in air with a chemiluminescence aerosol detector.
    Analytica Chimica Acta. Roč. 410, 1-2 (2000), s. 159-165. ISSN 0003-2670. E-ISSN 1873-4324
    Grant CEP: GA ČR GA203/98/0943
    Grant ostatní: COPERNICUS(BE) SUB-AERO EVK2-1999-000327
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 1.849, rok: 2000

    Modification of the luminol solution by means of addition of various complexones and surfactants has been investigated to eliminate interferences from gaseous co-pollutants in the determination of ambient nitrogen dioxide using a chemiluminescence aerosol detector. The simultaneous presence of EDTA and triton X-100 together with sulphite and iodide in the luminol solution suppressed interferences from ozone and peroxyacetyl nitrate to a negligible level and no scrubbers or corrections of the NO2 measurements were needed.