Počet záznamů: 1  

Irradiation setup at the U-120M cyclotron facility

  1. 1.
    0489398 - ÚJF 2019 RIV NL eng J - Článek v odborném periodiku
    Křížek, Filip - Ferencei, Jozef - Matlocha, Tomáš - Pospíšil, Jan - Príbeli, Peter - Raskina, Valentina - Isakov, Artem - Štursa, Jan - Vaňát, Tomáš - Vysoká, K.
    Irradiation setup at the U-120M cyclotron facility.
    Nuclear Instruments & Methods in Physics Research Section A. Roč. 894, č. 6 (2018), s. 87-95. ISSN 0168-9002. E-ISSN 1872-9576
    Grant CEP: GA MŠMT LM2015056; GA MŠMT LM2015058
    Institucionální podpora: RVO:61389005
    Klíčová slova: particle irradiation facility * radiation hardness * dosimetry * single event effect * single event upset * single event transient
    Obor OECD: Nuclear physics
    Impakt faktor: 1.433, rok: 2018

    This paper describes parameters of the proton beams provided by the U-120M cyclotron and the related irradiation setup at the open access irradiation facility at the Nuclear Physics Institute of the Czech Academy of Sciences. The facility is suitable for testing radiation hardness of various electronic components. The use of the setup is illustrated by a measurement of an error rate for errors caused by Single Event Transients in an SRAM-based Xilinx XC3S200 FPGA. This measurement provides an estimate of a possible occurrence of Single Event Transients. Data suggest that the variation of error rate of the Single Event Effects for different clock phase shifts is not significant enough to use clock phase alignment with the beam as a fault mitigation technique.
    Trvalý link: http://hdl.handle.net/11104/0283829

     
     