Počet záznamů: 1  

Two Methods of Mechanical Noise Reduction of Recorded Speech During Phonation in an MRI devic

  1. 1.
    0363068 - ÚT 2012 RIV SK eng J - Článek v odborném periodiku
    Přibil, J. - Horáček, Jaromír - Horák, Petr
    Two Methods of Mechanical Noise Reduction of Recorded Speech During Phonation in an MRI devic.
    Measurement Science Review. Roč. 11, č. 3 (2011), s. 92-98. ISSN 1335-8871. E-ISSN 1335-8871
    Grant CEP: GA ČR GA102/09/0989
    Výzkumný záměr: CEZ:AV0Z20760514; CEZ:AV0Z20670512
    Klíčová slova: speech processing * noise reduction * NMR imaging
    Kód oboru RIV: BI - Akustika a kmity
    Impakt faktor: 0.418, rok: 2011

    The paper presents two methods of noise reduction of speech signal recorded in an MRI device during phonation for the human vocal tract modelling. The applied approach of noise signal cleaning is based on cepstral speech analysis and synthesis because the noise is mainly produced by gradient coils, has a mechanical character, and can be processed in spectral domain. Our first noise reduction method is using real cepstrum limitation and clipping the “peaks” corresponding to the harmonic frequencies of mechanical noise. The second method is coming out from substation of the short-time spectra of two signals recorded withal: the first includes speech and noise, and the second consists of noise only. The resulting speech quality was compared by spectrogram and mean periodogram methods.
    Trvalý link: http://hdl.handle.net/11104/0199174

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.