    Tensile Stress Analysis of the Ceramic Head Endoprosthesis with different Micro.
    XII Mediterranean Conference on Medical and Biological Engineering and Computing 2010. Chalkidiki: Springer, 2010, s. 1-4. IFMBE Proceedings, 29. ISBN 978-3-642-13038-0. ISSN 1680-0737.
    [Mediterranean Conference on Medical and Biological Engineering and Computing MEDICON 2010 /12./. Chalkidiki (GR), 27.05.2010-30.05.2010]
    Výzkumný záměr: CEZ:AV0Z20760514
    Klíčová slova: hip joint endoprosthesis * shape deviations * reliability
    Kód oboru RIV: BO - Biofyzika

    The paper deals with the problems of ceramic head of hip joint endoprosthesis destructions in vivo, and with assessing the impact of shape deflections of conical surfaces on the tensile stress in the head. Concerned are shape deviations from the ideal conical surfaces of the stem and the head of the endoprosthesis. The shape deviations may be modelled at the macro-level (deviation from the nominal degree of taper), at the micro-level - when the stochastic distribution of unevenness on the contact areas is respected. The problem of stress in ceramic heads was solved using the finite element method – system ANSYS under ISO 7206-5 loading. There are analysed two different groups of micro shape deviations. The first is three variants of the sizes of the measured micro shape deviations (measured,doubled and halved). The second group of the analysed shapes deviations contained modeled deviations with the linear transformation of the measured deviations along the cone depth.
    Trvalý link: http://hdl.handle.net/11104/0192988

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.