Počet záznamů: 1
Spectrometry of the let LET with track etched detectors-correlation with proportional counter measured spectra
- 1.0185242 - UJF-V 20013061 RIV GB eng J - Článek v odborném periodiku
Spurný, František - Bottollier-Depois, J. F. - Vlček, Bohumil
Spectrometry of the let LET with track etched detectors-correlation with proportional counter measured spectra.
Radiation Measurements. Roč. 32, 1-6 (2001), s. 193-197. ISSN 1350-4487. E-ISSN 1879-0925
Grant CEP: GA ČR GA202/01/0710; GA AV ČR KSK1010104
Výzkumný záměr: CEZ:AV0Z1048901
Klíčová slova: experimental microdosimetry * track etched LET spectrometry * tissue equivalent proportional counter
Kód oboru RIV: BG - Jaderná, atomová a mol. fyzika, urychlovače
Impakt faktor: 1.003, rok: 2001
The work describes the results of comparison of two experimental microdosimetry methods: tissue equivalent proportional counter, and LET spectrometer based on the chemically etched PADC track detector. The methods were compared in the region of common sensibility between 10 and 700 keV/m. It is shown that the results obtained by both methods are similar, some differences are, nevertheless, observed.
Trvalý link: http://hdl.handle.net/11104/0081650
Počet záznamů: 1