Sol-gel column technology for capillary isoelectric focusing of microorganisms and biopolymers with UV or fluorometric detection
- 1.0162450 - UIACH-O 20033013 RIV DE eng J - Článek v odborném periodiku
Horká, Marie - Planeta, Josef - Růžička, F. - Šlais, Karel
Sol-gel column technology for capillary isoelectric focusing of microorganisms and biopolymers with UV or fluorometric detection.
Electrophoresis. Roč. 24, č. 9 (2003), s. 1383-1390. ISSN 0173-0835. E-ISSN 1522-2683
Grant CEP: GA AV ČR IAA4031901; GA AV ČR IAA4031302; GA ČR GA203/02/1447
Výzkumný záměr: CEZ:AV0Z4031919
Klíčová slova: sol-gel CIEF * microorganisms
Kód oboru RIV: CB - Analytická chemie, separace
Impakt faktor: 4.040, rok: 2003
The sol-gel surface modification is used for capillary isoelectric focusing of microorganisms and biopolymers. The coating strongly decreases the electroosmotic flow so that it enables the use of the short capillaries down to 100 mm in the separation length. The examples of capillary isoelectric focusing of the low-molecular-mass pI markers and mixed cultures of microbial populations of Escherichia coli, Candida albicans, Staphylococcus epidermidis and Enteroccocus faecalis with UV detection are shown. It is possible to quantify bacterial cells according their peak areas; the minimum detectable number of microbial cells was 5 x 102 1 x 103. The compatibility of sol-gel capillaries with sensitive fluorometric detection of fluorescent pI markers together with fluorescein labeled proteins is demonstrated.
Trvalý link: http://hdl.handle.net/11104/0059765