Low-conductivity background electrolytes in capillary zone electrophoresis - myth or reality?

  1. 1.
    0162187 - UIACH-O 200022 RIV DE eng J - Článek v odborném periodiku
    Horká, Marie - Šlais, Karel
    Low-conductivity background electrolytes in capillary zone electrophoresis - myth or reality?
    Electrophoresis. Roč. 21, č. 14 (2000), s. 2814-2827. ISSN 0173-0835. E-ISSN 1522-2683
    Grant CEP: GA AV ČR IAA4031901; GA AV ČR KSK2055603
    Kód oboru RIV: CB - Analytická chemie, separace
    Impakt faktor: 3.385, rok: 2000

    The parametr named relative velocity slope, SBGE,X, was introduced as a quantity characterising the peak broadening and the asymmetry. SBGE,X vs. analyte ionic mobility diagrams are suitable for the comparison of background electrolytes, BGEs, of different pKas and ionic mobility. It follows from both theoretical predictions of peak symmetry and their experimental verification that the optimum composition of BGEs is determined mostly by the suitable ionic mobility of the coion in dependence on the ionic mobility of the analyte.